Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 25: (1 year) CAL VER /5520 INSTRUMENT: Fluke 27: (1 year) CAL VER /5520 INSTRUMENT: Fluke 37: (1 year) CAL VER /5520 DATE: 26-May-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 152 CONFIGURATION: Fluke 5520A CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 25 27 service manual (PN 738138 Oct. 1984, Rev. 3, 11/94). # # Compatibility: # 5500/CAL 5.0 & MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R J F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature : 18 - 28 degC 1.004 DISP Humidity : <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DISPLAY TEST} 1.007 DISP Select VAC, verify that all display 1.007 DISP segments momentarily light up. 1.008 EVAL Did the UUT display test pass? 2.001 HEAD {AC VOLTAGE TESTS} 2.002 DISP Connect the 5520 to the UUT as follows: 2.002 DISP [32] 5520A NORMAL HI to UUT V Ohm Diode 2.002 DISP [32] 5520A NORMAL LO to UUT COM 2.003 HEAD AC VOLTAGE TESTS: {3.2V Range} 2.004 5520 3.2 2.700V 0.016U 100H SI 2W 3.001 HEAD AC VOLTAGE TESTS: {32V Range} 3.002 5520 32 27.00V 0.57U 10kH SI 2W 4.001 HEAD AC VOLTAGE TESTS: {320V Range} 4.002 5520 320 250.0V 5.3U 10kH SI 2W 5.001 HEAD AC VOLTAGE TESTS: {1000V Range} 5.002 5520 1000 1000V 33U 10kH SI 2W 6.001 HEAD {DC VOLTAGE TESTS} 6.002 DISP Select the DCV Function. 6.003 HEAD DC VOLTAGE TESTS: {3.2V Range} 6.004 5520 3.2 2.700V 0.004U 2W 7.001 HEAD DC VOLTAGE TESTS: {32V Range} 7.002 5520 32 27.00V 0.04U 2W 8.001 HEAD DC VOLTAGE TESTS: {320V Range} 8.002 5520 320 250.0V 0.4U 2W 9.001 HEAD DC VOLTAGE TESTS: {1000V Range} 9.002 5520 1000 1000V 2U 2W 10.001 HEAD {DIODE TEST} 10.002 M550 RNGLK 3.3V 10.003 DISP Select the DIODE Function. 10.004 5520 0.090V S 2W 10.005 EVAL Is the {beeper on}? 11.001 5520 0.090V 0.002U 2W 12.001 5520 0.110V S 2W 12.002 EVAL Is the {beeper off}? 13.001 5520 0.110V 0.002U 2W 14.001 EVAL Is the UUT displaying {OL}? 15.001 M550 * 15.002 ASK- U 15.003 HEAD {CONDUCTANCE TEST} 15.004 DISP Select the Ohms Function. 15.004 DISP 15.004 DISP Press RANGE twice to select nanosiemens. 15.005 HEAD CONDUCTANCE TEST: {32nS Range} 15.006 MATH MEM1 = 0 15.007 MEMI Use keyboard to enter actual UUT reading in nS: 15.008 MEME 15.009 MEMC nY 0.10U 16.001 ASK+ U 16.002 HEAD {RESISTANCE TESTS} 16.003 DISP Press RANGE to select Ohms. 16.004 DISP Press RANGE for 2s to select autorange. 16.005 HEAD RESISTANCE TESTS: {32 MOhm Range} 16.006 5520 32 10.00MZ 0.11U 2W 17.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 17.002 5520 3.2 1.000MZ 0.003U 2W 18.001 HEAD RESISTANCE TESTS: {320 kOhm Range} 18.002 5520 320 100.0kZ 0.3U 2W 19.001 DISP Connect the 5520 to the UUT as follows: 19.001 DISP [32] 5520A SENSE HI to UUT V Ohm Diode 19.001 DISP [32] 5520A SENSE LO to UUT COM 19.001 DISP [32] 5520A NORMAL HI to UUT V Ohm Diode 19.001 DISP [32] 5520A NORMAL LO to UUT COM 19.001 DISP 19.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 19.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 19.002 HEAD RESISTANCE TESTS: {32 kOhm Range} 19.003 5520 32 10.00kZ 0.03U CW 20.001 HEAD RESISTANCE TESTS: {3.2 kOhm Range} 20.002 5520 3.2 1.000kZ 0.003U CW 21.001 HEAD RESISTANCE TESTS: {320 Ohm Range} 21.002 5520 320 100.0Z 0.5U CW 22.001 DISP Disconnect the 5520A from the UUT. 22.002 HEAD {AC MICROAMP TEST} 22.003 DISP Connect the 5520 to the UUT as follows: 22.003 DISP [32] 5520A AUX HI to UUT mA/uA 22.003 DISP [32] 5520A AUX LO to UUT COM 22.004 DISP Select the AC uA Function. 22.005 HEAD AC MICROAMP TEST: {3200uA Range} 22.006 5520 3200 2000uA 32U 500H SI 2W 23.001 HEAD {AC MILLIAMP TESTS} 23.002 DISP Select the AC mA/A Function. 23.003 HEAD AC MILLIAMP TESTS: {32mA Range} 23.004 5520 32 27.00mA 0.42U 50H SI 2W 24.001 HEAD AC MILLIAMP TESTS: {320mA Range} 24.002 5520 320 200.0mA 3.2U 50H SI 2W 25.001 HEAD {DC MILLIAMP TEST} 25.002 DISP Select the DC mA/A Function. 25.003 HEAD DC MILLIAMP TEST: {32mA Range} 25.004 5520 32 27.00mA 0.22U 2W 26.001 HEAD {DC MICROAMP TEST} 26.002 DISP Select the DC uA Function. 26.003 HEAD DC MICROAMP TEST: {3200 uA Range} 26.004 5520 3200 2700uA 22U 2W 27.001 HEAD {DC AMP TEST} 27.002 DISP Select the DC mA/A Function. 27.003 DISP Connect the 5520 to the UUT as follows: 27.003 DISP [32] 5520A 20A to UUT A 27.003 DISP [32] 5520A AUX LO to UUT COM 27.004 HEAD DC AMP TEST: {10A Range} 27.005 5520 10 5.00A 0.06U HC 2W 28.001 HEAD {AC AMP TEST} 28.002 DISP Select the AC mA/A Function. 28.003 HEAD AC AMP TEST: {10A Range} 28.004 5520 10 5.00A 0.09U 1kH SI HC 2W 29.001 END